Spectroradiometer & Integrating Sphere System Spectroradiometer & Integrating Sphere System

Spectroradiometer & Integrating Sphere System

PCE Series Opto-Electric Measurement System

Simple Operation, Fast Speed, High Test Accuracy

This system delivers various photometric, colorimetric and electrical parameters as required by CIE standards.
Measurement parameters are luminous flux (with integrating sphere), relative spectral power distribution, chromaticity coordinate, CCT(correlated colour temperature), CRI (color rendering index), SDCM (standard deviation of color matching), color purity, peak wavelength, dominant wavelength, etc.

Features

  • Milliseconds Measurement Speed for LED
    Supporting LED special pulse DC power supply, it can realize the transient optical characteristics measurement (pulse measurement) and steady-state optical characteristics measurement (DC measurement) of LED, and realize the millisecond measurement speed and scientific measurement accuracy of LED under test.
  • Automatic Rise Time Determaination
    It can automatically describe the change curve of luminous flux/power/voltage/current with time after the light source is started, and can automatically judge the rise time and stable time of the light source (the criterion can be input by the user).
  • Comprehensive Measurement Functions
    With voltage, current, luminous flux, wavelength multi-grade classification and white LED color partition function, to meet the high requirements of LED production line fast sorting test.

Technical Specifications

ModelPCE Series Large Integrating Sphere SystemPCE Series Small Integrating Sphere System
Test SamplesLED modules, LED light sources, LED luminaires, etc.Individual LEDs and LED modules
SpectroradiometerIntegrated photometric, colorimetric, and electrical measurement solution built around a high-speed spectroradiometer (configured to order)
Wavelength Range380 nm–780 nm (other wavelength ranges available upon request)
Wavelength Accuracy0.3nm
Stray Light5E-5 (varies by order configuration)
Chromaticity Coordinate Accuracy0.0003 (under CIE Standard Illuminant A; varies by order configuration)
Luminous Flux Measurement Range0.01–200,000 lm (with different integrating spheres)
DC Power SupplyOptional pulsed or conventional DC power supply; range up to 300 V
AC Power Supply500 VA/1 kVA/3 kVA/6 kVA power supplies available (other configurations available upon request)
Digital Power MeterVarious power meter models available; accuracy class up to 0.15

HAAS-2000 High Accuracy Array Spectroradiometer (Laboratory Grade)

The Leading Speed Spectroradiometer

HAAS-2000 High Accuracy Array Spectroradiometer delivers various photometric, colorimetric and radiometric parameters as required by CIE standards.
Measurement parameters are luminous flux (with integrating sphere), relative spectral power distribution, chromaticity coordinate, CCT(correlated colour temperature), CRI (color rendering index), SDCM (standard deviation of color matching), color purity, peak wavelength, dominant wavelength, etc.

Features

  • Innumerable Awards & Honors
    "National High-Tech Program (863 program) Research Achievement", "US Patent Granted (No.: US 7,978,324 B2)", "China Excellent Patent Award", "Independent Innovation Product", "National Key& New Product",etc.
  • Remarkable Stray Light Control
    Thanks to the highly matched high-precision grating and science-grade TC-cooling detector, and using BWCT technology with modified NIST stray light correction technology,so the stray light reaches the extremely great level of 5.0E-05.
  • Wide Linear Dynamic Range
    The wide linearity dynamic range of the HAAS-2000 is further extended by rigorous optical matching and EVERFINE SBCT patented technology.
  • Fast Measurement Speed
    HAAS-2000 measurement functions including but not limited to constant-state optical characteristics.The transient optical characteristics can be easily captured through the synchronous trigger function even the transient pulse is less than microseconds.
  • High Accuracy
    The instrument is specially designed for high-accuracy applications. Through the improved design of TE-cooling array CCD and accurate optical grating, along with the precise optical system and electronic circuit, the instrument delivers high resolution, high sensitivity and high SNR.
  • High Repeatability & Stability
    Extremely high measurement repeatability and stability are achieved by TC-cooling technology that controls the influence of ambient temperature to the level of ±0.05 ° C, and no mechanical scanning mechanism.
  • High Sensitivity
    The EVERFINE designed high-precision grating and HAMAMATSU TE-cooling array detector make the sensitivity of the instrument great to 0.01mcd, which is in the world's leading level.

Technical Specifications

ModelHAAS-2000
Wavelength Range380 nm–780 nm (other wavelength ranges available with alternative configurations)
Wavelength Accuracy0.3nm
Chromaticity Coordinate Accuracy0.0015 (under standard color light); 0.0003 (under CIE Standard Illuminant A)
Chromaticity Coordinate Repeatability0.00015 in x, 0.0002 in y (temperature-stabilized blue LED)
Stray Light5E-5
Correlated Color Temperature (CCT) Range1000-100,000K
Integration Time9ms-60s

SPSR-2030 (200-3000nm) Super Pure Spectral Radiation Scanning Analyzer

Full-spectrum Scanning in the range of 200-3000 nm

SPSR-2030 adopts double-monochromator design, which can realize super-pure&fine full-spectrum scanning ( 200-3000nm), the test parameters including but not limited to spectral power distribution, peak wavelength, dominant wavelength, CCT (correlated colour temperature), CRI (color rendering index), SDCM (standard deviation of color matching), color purity, and etc.

Features

  • Pure & Fine Full-spectrum Scanning
    SPSR-2030 is an expert in the analysis of the full spectrum of 200nm ~ 3000nm, using a number of EVERFINE patented technologies.
  • High Accuracy & Resolution
    The unique coaxial double monochromator technology eliminates the scanning synchronization error of two gratings, and the system achieves extremely low stray light level and high wavelength accuracy.
  • High Repeatability & Stability
    Adopting a number of patented technologies on optical design, mechanisms, electronic circuits and application software,it delivers extremely high repeatability and stability.

Technical Specifications

ModelSPSR-2030
Spectral Measurement Range200nm-3000nm
Spectral Bandwidth (FWHM)Up to 2 nm
Wavelength AccuracyUp to 0.2 nm
Stray LightDown to 10⁻⁸
Optional FunctionsOptional accessories enable measurements of radiant flux, irradiance, radiance, and other parameters

ATA-500/1000 Auto-Temperatured LED Opto-Electronic Analyzer

Meets the Standard of IESNA LM-85

ATA-500/1000 controls the temperature of the LED, and the photochromatic-electric performance of the LED under different junction temperatures can be measured and analyzed in pulse mode or DC mode on patented software.

Features

  • Characteristics of LED at Different Case / Junction Temperature
    Using the world's leading spectroradiometer technology, it measures the photochromic comprehensive characteristics of single LED, integrated package COB, LED module and other products at different case temperature/junction temperature.
  • High Measurement Accuracy & Speed
    The cor components won“National High-Tech Program (863 program) Research Achievement”,”US Patent Granted (No.: US 7,978,324 B2)”,””China Excellent Patent Award“,”Independent Innovation Product” and “National Key & New Product”.
  • Pulse&Constant Measurement Mode
    With the EVERFINE special test power supply, the transient optical characteristics of LED (pulse measurement) and constant optical characteristics measurement (DC measurement)can be measured.

Technical Specifications

ModelATA-500ATA-1000
Wavelength Range380nm-780nm
Wavelength Accuracy0.3nm
Integration Time9ms~60s
Temperature-Controlled Fixture Control MethodAutomatic TEC Temperature Control
Temperature Control Range5 °C–105 °C (custom ranges available)
Maximum Power of LED Module Under Test10W100W
Maximum Dimensions of LED Module Under Testφ50mmφ100mm

PMS-2000 UV-VIS-near IR Spectrophotocolorimeter (Scientific Grade)

Double-monochromator Based Spectroradiometer with Extremely High Accuracy

PMS-2000, with extremely low stray light and greatly high wavelength accuracy, is trusted by high-end laboratories and scientific research institutions. It not only realizes the high-accuracy test of various light sources and fluorescent materials such as LED, energy-saving lamps, halogen lamps, but also be applied to various professional measurement systems, which are widely used in the fields of photobiosafety, medical devices, ultraviolet, infrared, printing, coatings and so on.

Features

  • Perfect Stray Light Control and Wavelength Accuracy
    The unique coaxial double monochromator technology eliminates the scanning synchronization error of two gratings, and the system achieves extremely low stray light level and high wavelength accuracy.
  • High Measurement Accuracy
    Adopting a number of patented technologies, its accurate value can trace standard to the US NIST or China NIM.
  • Powerful Software & Friendly UI
    It has a high degree of intelligence, powerful software functions and complete test parameters.
  • Self-calibration & Inspection Function
    The calibration and inspection work is very simple, and the working condition of the instrument can be recorded synchronously in order to monitor whether the instrument is working in the best state.

Technical Specifications

ModelPMS-2000
Wavelength Range200nm-800nm
Wavelength AccuracyUp to ±0.1 nm
Wavelength Repeatability0.1nm
Spectral Sampling IntervalSelectable: 0.1 nm, 1 nm, or 5 nm
Luminous Flux Measurement Range0.01 lm–1.9999 × 10⁵ lm (with different integrating spheres)
Chromaticity Coordinate Accuracy0.0003 (relative to a standard light source with stability better than 0.0001 and traceable to NIM)
Stray Light10E-8

HAAS-1200 Accurate Array Spectroradiometer (Advanced industry grade)

Ideal Product to Replace General Mechanical Scanning Spectrometers

HAAS-1200 quickly and accurately measures the spectral power distribution in the range of 380nm-1000nm, and achieves the photometric, colorimetric and radiometric parameters of various luminairs and lighting sources according to the requirements of the International Commission on Lighting (CIE) standard publications.

Features

  • High Speed & Stability & Cost-performance
    Advanced industrial products based on HAAS-2000 core technology, which has the performance of high speed and high stability.
  • High Sensitivity, High SRN, Wide LDR
    Using international high-quality grating and high-performance array CCD, the array CCD has high sensitivity, low noise and wide linearity dynamic range, referring to CIE's requirements for light and color measurement.
  • High Matching Degree of Optical System,Pefect Stray Light Control
    Adopting a number of patented technologies,the optical matching between grating and array detector is high and the influence of stray light is minimal.

Technical Specifications

ModelHAAS-1200
Wavelength Range380nm~1000nm
Wavelength Accuracy0.3nm
Spectral Bandwidth (FWHM)3.0nm
Chromaticity Coordinate Accuracy0.003 (under standard color light)
Spectral Linearity1.00%
Integration Time9ms~100s
Luminous Flux Measurement Range0.01–200,000 lm (with different integrating spheres)
Photometric AccuracyClass 1

CAS-200 Compact Array Spectrometer

Fast and accurate measurement in the range from 380nm to 780nm

CAS-200 can quickly and accurately measure the spectral power distribution in the range of 380nm ~ 780nm, mainly measuring: the relative spectral power distribution, chromaticity coordinate, color temperature, color rendering index, color tolerance, peak wavelength, FWHM, luminous flux (with integrating sphere) of traditional lamps/light sources, single LED, LED modules and other electric light sources.

Features

  • Fast Measurement Speed
    Using precision array CCD and high quality grating, the electrical signal is formed after precision optical splitting system and high performance array detector, and the test speed is fast.
  • High Matching Design with Low Stray Light
    The high quality grating and the optical system of the array detector have high matching degree, small influence of stray light, and the test accuracy meets the requirements.

Technical Specifications

ModelCAS-200
Wavelength Range380nm~780nm
Wavelength Accuracy0.3nm
Spectral Bandwidth (FWHM)3.0nm
Integration Time1ms~10s
Spectral Linearity1.5%
Luminous Flux Measurement Range0.01–200,000 lm (with different integrating spheres)